Transducer Protector-FT0211

Products

Item-Name:

Transducer Protector-FT0211

Specification:

Model:FT0211

Dimeter:23mm

Membrane:PTFE

Pore Size:0.2um

Filtration area:2.5cm²

Package:5000pcs/PK

More-Details:

◆ Introduction

Transducer protectors are used in hemodialysis blood lines to keep the blood side of the circuit separated from the machine side and to prevent  contamination of the machine by the blood flowing through the circuit. This contamination could be very dangerous and can lead to patient cross contamination with hepatitis B or other virus

 

◆ Features

1. Hydrophobic membrane with 0.2 micron pore size.

2. Prevents contamination of the internal pressure monitoring lines.

3. Antibacterial hydrophobic air filter with female  luer lock/male luer lock

4. Connects blood tubing with dialysis machine

5. Protective hydrophobic barrier allows only sterile air to pass through, protecting patients
    and equipment from cross contamination.
6. Prevents blood and dialyzing fluid from damaging sensitive transducer monitoring devices

 

Transducer Protector commissioned SGS test 

► Approval of biocompatibility test 

Cytotoxicity test (direct contac, indirect contact)

Heat source test

Skin allergy test 

Intradermal irritation test 

Toxicity test for Acute systemic 

Hemolysis test (direct contact, indirect contact)

 
◆ Specification

  Effective Filtration Area

  2.5 cm2

  Materials of Construction Media

  PTFE LAMINATE (hydrophobic membrane)

  Inlet/Outlet Connections

  Inlet: MLL;Luers conform to ISO594

  Outlet: FLL;Luers conform to ISO594

  Sterilization Compatibility

  E.T.O.

  Maximum Operating Temperature

 60°C

  Pore Size 

  0.2 µm

  Biological Safety 

  Materials of construction pass USPClass VI



◆Design Patent-Transducer Protector
03-1TP新式樣專利(D 130528)